Comprehensive measurement functions
4230 LCR Meter not only measures all the parameters normally encountered
in a standard LCR meter but it measures many more in addition to give
a truly versatile instrument.
Standard parameters
Impedance - Z
Series and Parallel Resistance - R
Capacitance - C
Inductance - L
Dissipation Factor - D
Quality Factor - Q
Phase Angle – Ø
Additional parameters
Reactance – X
Conductance – G
Susceptance – B
DC Resistance – Rdc (simulated measurement
using 5 Hz square wave)
Flexible full remote control
Full control and output capability is a significant benefit of the
4230 LCR Meter. Three interfaces are included as standard: RS232,
GPIB and Bin Handler. Remote control and printing are both very
straightforward.
Fast, accurate repeatable measurements
Testing electronic components is easy with the 4230 LCR Meter.
Connect the component and the meter accurately measures the value -
displaying the selected primary and secondary value with a basic accuracy
of 0.1%.
Up to 127 instrument set-up conditions can be saved and recalled
for rapid and consistent measurements.
Connection to the component to be tested is made using one of a range
of 4 terminal Kelvin connectors.
Full adjustable range - 100 Hz to 200 kHz
With a variable frequency from 100 Hz to 200 kHz components can be
tested at the correct operating frequency.
Superb performance and economical price
The 4230 LCR Meter provides unbeatable performance, 100 Hz to 200
kHz frequency range, 0.1% basic accuracy and comprehensive measurement
functions.
RS232, GPIB and bin handler interfaces are all in the instrument
as standard.
Accessories
Optional accessories include a four terminal component fixture, Kelvin
clips, SMD Tweezer and BNC to 4 terminal component fixture.
Technical specifications
Basic accuracy
0.1%
Test conditions
Frequency range
100 Hz to 200 kHz
Steps
390
Drive level (AC measurements)
50 mV, 100 mV, 250 mV and 1 Vrms
Drive level (Rdc)
0.5 V (simulated measurement using 5 Hz square wave)
Bias voltage
2 V DC
Measurement ranges
|Z| R X 0.01 mΩ to
100.0 MΩ
|Z| G B 0.0001 nS to 1000.0
S
Cs Cp
0.01 pF to 1 F
Ls Lp
0.01 µH to 100 kH
D
0.0001 to 9.9999
Q
0.1 to 9999.9
Ø
-180° to +180°
Δ
-999.99% to 999.99%
DCR
0.01 mΩ to 100.00 MΩ, simulated Rdc measurement using 5 Hz square wave
Measurement speed
Fast 5 measurements per second
Medium 3 measurements per second
Slow 2 measurements
per second
Comparator test mode
Upper and lower test limits in units and percentage
Ranging
Auto and Hold
Trigger
Internal, Manual, External and Bus
Display and indicators
240 x 64 pixel dot matrix LCD
PASS, FAIL and BIAS LED indicators
Internal loudspeaker
Connections
4-terminal BNC guarded connectors
Memory
127 instrument set-up conditions can be saved
Interfaces
RS232 port, Handler Interface, GPIB
Power supply
Input voltage 90 to 132 V AC, 198 to
264 V AC (selectable)
Frequency 47 to 66 Hz
Power 45 VA typical
Environmental
Installation category II in accordance with IEC664
Temperature range Operating: 10°C to 40°C
Relative humidity Up
to 90% non-condensing
Pollution degree
2 (mainly non-conductive)
Altitude Up to 2000 m
Safety
Complies with the requirements of EN61010-1
EMC
Immunity EN61326-1
Emissions
EN55022 Class A
Mechanical
Height
150 mm
Width
320 mm
Depth
300 mm
Weight 6 kg without accessories
Order codes and options
4230 LCR Meter
Supplied with user manual, power cord and Kelvin leads (AC-6870K1) |
|
1J4230 |
|
Accessories
|
| Kelvin clips (fine
jaws). |
|
1EVA40100 |
| Kelvin clips (large
jaws) |
1EVA40180 |
| BNC to 4-terminal
component fixture |
1EV1006 |
| 4-terminal component
fixture |
FX-0000C1 |
| SMD Tweezers
|
1EVA40120 |
|